Theory: Alpha Particles in Matter
Why Alpha Range Matters
An alpha particle is a helium-4 nucleus emitted with a single, well-defined energy, typically 4 to 9 MeV. Because it is heavy and doubly charged, it ionizes densely and loses its energy in a short, nearly straight track. Its range sets the design of alpha detectors (the dead layer of a silicon detector, the window of a proportional counter), the thickness of sources and sample deposits in alpha spectrometry, and radiation protection: alphas cannot pass the dead outer layer of skin, but an alpha emitter taken into the body deposits all its energy in a few cells. Smoke detectors, static eliminators, and radon dosimetry all rely on this behaviour.
Stopping Power and the Bragg Curve
The rate of energy loss, the stopping power −dE/dx, is described by the Bethe formula. For a heavy charged particle it rises as the particle slows (roughly as 1/E), so most of the energy is deposited near the end of the track. A plot of −dE/dx against depth is the Bragg curve; its sharp maximum just before the particle stops is the Bragg peak, the same effect used in proton therapy. Because every alpha from a source starts with the same energy, all of them stop at nearly the same depth, with only a small spread called range straggling.
Range in Air: the Geiger Rule
Inverse: E = (R / 0.318)^(2/3)
At pressure P: R_P = R₀ × (101.3 kPa / P)
The range depends on the number of atoms along the path, so it is inversely proportional to the gas density. At constant temperature, halving the pressure doubles the range. Measuring the count rate while the pressure is changed is a classic way to find the range: the count drops to background once the detector is beyond the range. A path of x cm at pressure P is equivalent to x × P / 101.3 cm at STP.
Range in Solids: the Bragg–Kleeman Rule
Thickness (µm) = 10 × R_m / ρ (ρ in g/cm³)
Quoting the range as a mass range (density × thickness) makes it nearly the same for all materials; it rises only slowly with atomic mass number. A 5.3 MeV alpha from Po-210 is stopped by about 24 µm of aluminium, under 7 µm of gold, or about 45 µm of tissue, less than the 50–100 µm dead outer layer of skin. A foil thinner than the stopping thickness lets the alphas through with reduced energy; converting the foil to an air-equivalent path and applying the Geiger rule gives the energy that emerges. These rules are empirical approximations with an accuracy of a few percent; precise work uses tabulated stopping powers such as the NIST ASTAR data.
Range–Energy Curves, Stopping Power, and Straggling
A range–energy curve gives the range of an alpha of any energy in one look. Its inverse slope is the stopping power: S = dE/dx = 1/(dR/dE). For R = 0.318 E^1.5, dR/dE = 0.477 E^0.5, so the stopping power grows as the alpha slows down. That is why energy loss peaks near the end of the track.
Individual alphas do not all stop at exactly the same depth. Random collisions spread the ranges around the mean range R̄ with a nearly normal distribution of width σ, about 1% of R̄ for alphas in air. A transmission curve, the fraction of alphas that still reach a detector after a path x, therefore falls smoothly: T(x) = ½ erfc[(x − R̄)/(√2 σ)]. Its 50% point is the mean range, and the tangent at that point meets the axis at the extrapolated range, R̄ + √(π/2) σ.
Apparatus
The equipment used to measure alpha ranges in gases and foils. In the simulation these are modelled for you.
Instructions: Running the Virtual Experiment
You are the physicist characterising alpha sources and absorbers. The simulation has two immersive 3D work areas and two graph work areas. The parts build from undergraduate work in Parts A and B to graduate-level analysis in the last step of Parts C and D; the level of each part is shown beside its title. In each one you set the apparatus, work out the answer yourself, enter it, and press Check. Record every reading and calculation in your lab report tables (modelled in the Example Report), with a labelled screenshot of each setup. Drag in any 3D view to orbit, and scroll to zoom.
Simulation: Alpha Range Chamber, Foil Absorber Bench, and Range Graphs
Alpha particles lose energy steadily in a gas and stop at a well-defined range. The range depends on the alpha energy and on the gas density, so it changes with pressure. Part A uses this work area.
Drag to orbit · scroll to zoom. 3D range chamber: alpha tracks leave the source and stop at their range in the gas. The detector moves with your distance setting, and lowering the pressure makes the tracks longer.
Setup
Readings
Geiger rule (air at STP, 101.3 kPa): R = 0.318 E^1.5, with R in cm and E in MeV. The range is inversely proportional to the gas pressure.
Work it out, then check
| Source | E (MeV) | x (cm) | P (kPa) | R₀ (cm) | R_P (cm) | E at detector (MeV) | Counts / 60 s |
|---|
In a solid, the range is quoted as a mass range in mg/cm², which depends only weakly on the material. Dividing by the density gives the thickness that stops the alphas. Part B uses this work area.
Drag to orbit · scroll to zoom. 3D foil bench: alphas pass through the foil, or stop in it, and the wall chart plots the Bragg curve of energy loss against depth in the chosen material, with your foil thickness marked.
Setup
Readings
Bragg–Kleeman rule: R_m (mg/cm²) = 0.56 A^(1/3) R_air (cm). Thickness (µm) = 10 × R_m / ρ, with ρ in g/cm³.
Work it out, then check
| Source | Material | R_m (mg/cm²) | Stopping thickness (µm) | Foil (µm) | E out (MeV) | ΔE (MeV) |
|---|
Range–energy curves let you read the range of an alpha of any energy, or its energy from a measured range, in one step. Part C uses this work area.
Range–energy curve for alphas in the chosen material, with the four laboratory sources marked and your alpha energy highlighted.
Show the plotted values
Setup
Readings
R_air = 0.318 E^1.5 (cm, MeV); t = 10 × 0.56 A^(1/3) × R_air / ρ (µm). Simplified rules, best between about 4 and 8 MeV.
Work it out, then check
| Material | E (MeV) | Range | Measured air range (cm) | E from range (MeV) | Stopping power |
|---|
Counting alphas through a growing air path gives the transmission curve. Its 50% point is the mean range, and straggling makes the fall-off gradual instead of a sharp step. Part D uses this work area.
Integral range (transmission) curve in air at STP: the fraction of alphas that reach a detector at each distance. Straggling rounds the end of the curve; the dashed tangent at the 50% point meets the axis at the extrapolated range.
Show the plotted values
Setup
Readings
T(x) = ½ erfc[(x − R̄)/(√2 σ)], with R̄ from the Geiger rule and σ = 1.2% of R̄.
Work it out, then check
| Source | R̄ (cm) | σ (cm) | Path x (cm) | Fraction transmitted | R_ex (cm) | x for T = 10⁻⁴ (cm) |
|---|
These immersive 3D work areas and interactive graph work areas are the main version; the accessible copy delivers the same set-up, work-out, and check exercises with native controls and screen-reader announcements.
Team Questions
Example Lab Report
Sample report showing the expected format and level of detail. Use it as a guide for your own submission, and include labelled screenshots of each setup. Counts are random, so your counts will differ slightly.
Alpha Particles: Range, Energy Loss, and Stopping Power
Nuclear Physics | Section: [Your Section] | Date: [Date]
Lab Members: [Names of all members present]
Purpose
To find the range of alpha particles in air and its dependence on pressure, the energy remaining after a given path, and the mass range, stopping thickness, and energy loss in metal and tissue foils, and to use range–energy and transmission curves to find stopping power, straggling, and the path needed to stop alphas completely.
Theory
The Geiger rule relates range in air to energy; range scales inversely with pressure; the Bragg–Kleeman rule converts air range to mass range in another material.
Table A: Range Chamber Undergraduate
| Source | E (MeV) | x (cm) | P (kPa) | R₀ (cm) | R_P (cm) | E at detector (MeV) | Counts / 60 s |
|---|---|---|---|---|---|---|---|
| Po-210 | 5.30 | 2.0 | 101.3 | 3.88 | 3.88 | 3.27 | 476 |
| Po-210 | 5.30 | 5.0 | 101.3 | 3.88 | 3.88 | 0 (stopped) | 1 |
| Po-210 | 5.30 | 3.0 | 50.0 | 3.88 | 7.86 | 3.85 | 238 |
| Am-241 | 5.49 | 2.0 | 101.3 | 4.09 | 4.09 | 3.51 | 484 |
| Cm-244 | 5.80 | 6.0 | 70.0 | 4.44 | 6.43 | 0.95 | 63 |
| Po-212 | 8.78 | 6.0 | 101.3 | 8.27 | 8.27 | 3.71 | 67 |
Cm-244 at 6.0 cm: count drops to background above 101.3 × 4.44/6.0 = 75.0 kPa.
Table B: Foil Absorbers Undergraduate
| Source | Material | R_m (mg/cm²) | Stopping thickness (µm) | Foil (µm) | E out (MeV) | ΔE (MeV) |
|---|---|---|---|---|---|---|
| Po-210 | Aluminium | 6.52 | 24.1 | 5.0 | 4.54 | 0.76 |
| Po-210 | Copper | 8.67 | 9.7 | 5.0 | 3.26 | 2.04 |
| Po-210 | Gold | 12.64 | 6.6 | 5.0 | 2.03 | 3.27 |
| Po-210 | Soft tissue | 4.50 | 45.0 | 30.0 | 2.55 | 2.75 |
| Am-241 | Gold | 13.33 | 6.9 | 5.0 | 2.33 | 3.16 |
| Po-212 | Soft tissue | 9.60 | 96.0 | 50.0 | 5.38 | 3.40 |
Table C: Range–Energy Curves Upper undergraduateGraduate step
| Material | E (MeV) | Range | Measured air range (cm) | E from range (MeV) | Stopping power |
|---|---|---|---|---|---|
| Air (STP) | 5.30 | 3.88 cm | 4.0 | 5.41 | 0.911 MeV/cm |
| Air (STP) | 8.78 | 8.27 cm | 8.0 | 8.59 | 0.708 MeV/cm |
| Aluminium | 5.30 | 24.14 µm | 4.0 | 5.41 | 146.4 keV/µm |
| Aluminium | 8.78 | 51.48 µm | 8.0 | 8.59 | 113.7 keV/µm |
| Soft tissue | 5.30 | 45.03 µm | 4.0 | 5.41 | 78.5 keV/µm |
| Soft tissue | 8.78 | 96.01 µm | 8.0 | 8.59 | 61.0 keV/µm |
Graduate step: S = 1/(dR/dE). In air S rises from 0.911 MeV/cm at 5.30 MeV to 1.482 MeV/cm at 2.00 MeV; in aluminium from 146.4 to 238.2 keV/µm. Include a screenshot of one range–energy curve with your alpha marked.
Table D: Transmission Curve in Air (σ = 1.2% of R̄) Upper undergraduateGraduate step
| Source | R̄ (cm) | σ (cm) | Path x (cm) | Fraction transmitted | R_ex (cm) | x for T = 10⁻⁴ (cm) |
|---|---|---|---|---|---|---|
| Po-210 | 3.880 | 0.0466 | 3.83 | 0.859 | 3.938 | 4.053 |
| Po-210 | 3.880 | 0.0466 | 3.93 | 0.142 | 3.938 | 4.053 |
| Am-241 | 4.091 | 0.0491 | 4.05 | 0.796 | 4.152 | 4.273 |
| Cm-244 | 4.442 | 0.0533 | 4.50 | 0.138 | 4.509 | 4.640 |
| Po-212 | 8.273 | 0.0993 | 8.20 | 0.769 | 8.398 | 8.642 |
Graduate step: T = 10⁻⁴ needs z = 3.719 standard deviations beyond R̄, so x = R̄ + 3.719σ: only 0.17 cm past the mean range for Po-210 and 0.37 cm for Po-212.
Discussion
The count stayed high until the detector passed the range and then fell abruptly to background, confirming that alphas from one source have a nearly single range. Halving the pressure doubled the range, as expected for a range set by the number of atoms along the path. The energy at the detector fell faster than the distance travelled because the stopping power rises as the alpha slows: at 6 cm and 70 kPa the Cm-244 alphas had used 93% of their range but arrived with 0.95 MeV.
The mass ranges grew only from 4.5 to 12.6 mg/cm² between tissue and gold, while the stopping thicknesses fell from 45 to 6.6 µm because of the density. In the thin foils the energy lost was much more than the fraction of the range crossed, which is the Bragg curve at work. Po-210 alphas cannot penetrate the dead layer of skin, but the 8.78 MeV alphas of Po-212 could reach about 96 µm, near the living cells, which is one reason radon progeny matter for dose.
The range–energy curves showed the same alpha travelling centimetres in air but only tens of micrometres in solids, and the tissue range for Po-210 (45 µm) is inside the 50–100 µm dead layer of skin. Working backwards from a measured air range gave the alpha energy to within the rule's accuracy, which is how unknown alpha sources were first identified.
The graduate steps explained two features of the Bragg curve. Differentiating the range–energy rule showed the stopping power rising by about 60% between 5.3 and 2.0 MeV, so energy deposition climbs toward the end of the track. The transmission curve showed that straggling spreads the stopping points by only about 1% of the range: a path 3.7σ beyond the mean range, a fraction of a millimetre to a few millimetres of air, stops all but 1 alpha in 10⁴.
Conclusion
Alpha range follows the Geiger and Bragg–Kleeman rules within a few percent: a few centimetres of air or tens of micrometres of solid stops them, and most of their energy is deposited at the end of the track.
Practice Questions
Show all work and include units.