Theory: Neutron Sources and Gas Detectors
Why Neutron Detection Matters
Neutron detectors are used wherever neutrons are produced: reactor startup channels, health physics survey meters, well-logging tools in the oil industry, border portal monitors that look for plutonium, and nuclear safeguards. A neutron has no charge, so it passes through a gas without ionizing it. To detect it, the detector must first make it take part in a nuclear reaction that releases charged particles, and those particles ionize the gas. Technicians check these instruments against calibrated neutron sources, which is the first half of this lab.
Neutron Sources
| Source | How it makes neutrons | Yield used here | Half-life |
|---|---|---|---|
| AmBe | Am-241 alphas strike beryllium: ⁹Be(α,n)¹²C | 2.2 × 10⁶ n/s per Ci | 432 y |
| PuBe | Pu-239 alphas strike beryllium: ⁹Be(α,n)¹²C | 1.7 × 10⁶ n/s per Ci | 24 110 y |
| Cf-252 | Spontaneous fission (3.1% of decays), about 3.8 neutrons per fission | 2.3 × 10⁶ n/s per µg | 2.645 y |
φ = S/(4πr²) (point source, no scatter)
Count rate = φ × A × ε
Source neutrons are fast (a few MeV). The counter in Part A is wrapped in polyethylene, which slows them to thermal energies where the gas cross sections are large. Its overall efficiency for fast neutrons is only about 12%.
Gas Proportional Detectors
| Gas | Reaction | Q-value | Thermal σ |
|---|---|---|---|
| BF₃ (96% B-10) | ¹⁰B + n → ⁷Li + α | 2.79 MeV (6%); 2.31 MeV (94%, ⁷Li* then 0.48 MeV γ escapes) | 3840 b |
| He-3 | ³He + n → ³H + p | 0.764 MeV | 5330 b |
σ(E) = σ₀ √(0.025 eV / E) (the 1/v law)
Σ = nσ · ε = 1 − e^(−ΣL)
He-3 has the larger cross section and can be filled to several atmospheres, so it is more efficient. BF₃ is cheaper and rejects gamma rays better because its pulses are larger, but it performs poorly above about 1 atm. He-3 is now scarce, because it comes mainly from the decay of tritium.
The Pulse-Height Spectrum and the Wall Effect
When both reaction products stop in the gas, the pulse corresponds to the full Q-value, giving the full-energy peak. Near the tube wall one product may strike the wall before it deposits its energy. The products fly apart with equal and opposite momentum, so they share Q in inverse proportion to their masses. For BF₃ (Q = 2.31 MeV) the α takes 1.47 MeV and the ⁷Li takes 0.84 MeV. For He-3 the proton takes 0.573 MeV and the triton 0.191 MeV. The wall-effect continuum runs from the smaller share up to the full peak, with a step at the larger share. A discriminator is set below the lower edge, so nearly every neutron is counted while gamma pulses and noise are rejected.
Efficiency Curves, Moderators, and Room Return
Because the absorber cross section falls as 1/v, a bare tube's intrinsic efficiency ε(E) = 1 − exp[−Σ₀√(0.025/E)L] is high for thermal neutrons and drops steeply above a fraction of an electronvolt. Fast-neutron instruments therefore surround the tube with polyethylene: a thicker layer thermalises more neutrons but also absorbs or scatters away more of them, so the efficiency passes through a maximum. In a calibration, neutrons scattered from walls and floor add a room return B that hardly changes with distance, so the measured rate is C(r) = SAε/(4πr²) + B. B can be measured with a shadow cone that blocks the direct beam, or found by fitting C against 1/r².
Apparatus
The sources and detectors used to calibrate neutron instruments. In the simulation these are modelled for you.
Instructions: Running the Virtual Experiment
You are the detector technician calibrating neutron instruments. The simulation has two immersive 3D work areas and two graph work areas. The parts build from undergraduate work in Parts A and B to graduate-level analysis in the last step of Parts C and D; the level of each part is shown beside its title. In each one you set the conditions, work out the answer yourself, enter it, and press Check. Record every reading and calculation in your lab report tables (modelled in the Example Report), with a labelled screenshot of each setup. Drag in any 3D view to orbit, and scroll to zoom.
Simulation: Neutron Source Bench, Gas Detector Bench, and Calibration Graphs
A sealed neutron source emits neutrons equally in all directions. The flux at a distance r is the emission rate spread over a sphere of area 4πr². A moderated counter turns part of that flux into counts. Part A uses this work area.
Setup
Readings
S = yield × amount (× 2^(−t/T½) for Cf-252) · φ = S/(4πr²) · count rate = φ × A × ε. Room scatter is neglected.
Work it out, then check
| Source | Amount | Age (y) | r (cm) | S (n/s) | φ (n/cm²·s) | Predicted (cps) | Measured 60 s (cps) |
|---|
A gas proportional tube detects neutrons through a capture reaction that releases charged particles. The chance that a neutron is captured on its way along the tube depends on how many absorber atoms it passes. Part B uses this work area.
Setup
Readings
n = P/(kT), with k = 1.381 × 10⁻²³ J/K and T = 293 K · Σ = nσ · ε = 1 − e^(−ΣL) · σ(E) = σ₀ √(0.025 eV / E).
Work it out, then check
| Gas | P (atm) | L (cm) | E (eV) | n (cm⁻³) | Σ (cm⁻¹) | ε | Measured 60 s ε |
|---|
A bare gas tube counts thermal neutrons well and fast neutrons poorly. Plotting efficiency against energy shows why, and a moderator design study shows how a counter is made to respond to fast neutrons. Part C uses this work area.
Setup
Readings
ε(E) = 1 − exp[−Σ₀√(0.025/E)·L]. Moderator: f_th = 1 − e^(−t/2.5 cm), transmission T = e^(−t/7.0 cm), ε_mod = f_th·T·ε₀. Simplified training model.
Work it out, then check
| Gas | P (atm) | L (cm) | E (eV) | ε(E) | E at ε₀/2 (eV) | t (cm) | ε_mod |
|---|
Neutrons that scatter off walls and floor reach the detector from every direction, so a calibration must separate the source signal from room return. This survey uses a 1.0 Ci AmBe source (2.2 × 10⁶ n/s) and the 50 cm², ε = 0.12 counter. Part D uses this work area.
Setup
Readings
C(r) = SAε/(4πr²) + B, where B is the room return. Simplified training data; the measurements include a small random error.
Work it out, then check
| Room | r (cm) | Measured (cps) | Source-only (cps) | Room return (cps) | r where B = source (cm) | Fitted B (cps) |
|---|
This accessible version provides the same set-up, work-out, and check exercises as the 3D and graph work areas, using native controls, data tables, and screen-reader announcements.
Team Questions
Example Lab Report
Sample report showing the expected format and level of detail. Use it as a guide for your own submission, and include labelled screenshots of each setup. Your 60-second counts are random, so your measured values will differ slightly.
Neutron Sources and Gas Detectors: BF₃ and He-3
Nuclear Physics | Section: [Your Section] | Date: [Date]
Lab Members: [Names of all members present]
Purpose
To find the emission rate of three types of neutron source, the flux and count rate they give at a detector, and the efficiency and pulse-height spectrum of BF₃ and He-3 proportional tubes; then to use efficiency curves to design a moderated counter and to separate source signal from room return in a calibration survey, including a graduate-level optimisation and least-squares fit.
Theory
A point source spreads its neutrons over a sphere, so the flux falls as 1/r². A gas tube captures neutrons with probability 1 − e^(−ΣL), where Σ = nσ and σ follows the 1/v law.
Table A: Source Bench (face area 50 cm², ε = 0.12) Undergraduate
| Source | Amount | Age (y) | r (cm) | S (n/s) | φ (n/cm²·s) | Predicted (cps) | Measured 60 s (cps) |
|---|---|---|---|---|---|---|---|
| AmBe | 1.0 Ci | — | 50 | 2.20 × 10⁶ | 70.03 | 420.2 | 418.6 |
| AmBe | 1.0 Ci | — | 100 | 2.20 × 10⁶ | 17.51 | 105.0 | 106.3 |
| Cf-252 | 10 µg | 0 | 100 | 2.30 × 10⁷ | 183.0 | 1098 | 1101 |
| Cf-252 | 10 µg | 2.6 | 100 | 1.164 × 10⁷ | 92.60 | 555.6 | 553.9 |
| PuBe | 2.0 Ci | — | 100 | 3.40 × 10⁶ | 27.06 | 162.3 | 160.8 |
Doubling r from 50 to 100 cm cut the count rate by a factor of 4.00, as the inverse-square law predicts.
Table B: Gas Detector Bench (T = 293 K, beam 1000 n/s) Undergraduate
| Gas | P (atm) | L (cm) | E (eV) | n (cm⁻³) | Σ (cm⁻¹) | ε | Measured 60 s ε |
|---|---|---|---|---|---|---|---|
| BF₃ | 1.00 | 20 | 0.025 | 2.405 × 10¹⁹ | 0.09234 | 0.8422 | 0.8436 |
| BF₃ | 0.50 | 20 | 0.025 | 1.202 × 10¹⁹ | 0.04617 | 0.6028 | 0.6011 |
| BF₃ | 1.00 | 20 | 1 | 2.405 × 10¹⁹ | 0.01460 | 0.2532 | 0.2541 |
| He-3 | 4.00 | 5 | 0.025 | 1.002 × 10²⁰ | 0.5340 | 0.9308 | 0.9302 |
| He-3 | 4.00 | 5 | 1 | 1.002 × 10²⁰ | 0.08444 | 0.3444 | 0.3459 |
| He-3 | 1.00 | 5 | 0.025 | 2.505 × 10¹⁹ | 0.1335 | 0.4870 | 0.4851 |
Spectra: BF₃ main peak 2.31 MeV (small peak 2.79 MeV), wall-effect edges 0.84 and 1.47 MeV. He-3 peak 0.764 MeV, edges 0.191 and 0.573 MeV.
Table C: Efficiency Across Energy and Moderated Efficiency Upper undergraduateGraduate step
| Gas | P (atm) | L (cm) | E (eV) | ε(E) | E at ε₀/2 (eV) | t (cm) | ε_mod |
|---|---|---|---|---|---|---|---|
| BF₃ | 1 | 20 | 1 | 0.2532 | 0.2853 | 3.0 | 0.3834 |
| BF₃ | 1 | 20 | 10 | 0.0882 | 0.2853 | — | — |
| BF₃ | 2 | 30 | 1 | 0.5835 | 1.615 | — | — |
| He-3 | 4 | 5 | 0.025 | 0.9308 | 0.4545 | 3.0 | 0.4237 |
| He-3 | 4 | 5 | 1 | 0.3444 | 0.4545 | 6.0 | 0.3592 |
| He-3 | 1 | 10 | 0.1 | 0.4870 | 0.2110 | — | — |
ε(E) = 1 − exp[−Σ₀√(0.025/E)L]; E½ = 0.025[Σ₀L/(−ln(1 − ε₀/2))]²; ε_mod = (1 − e^(−t/2.5))e^(−t/7.0)ε₀.
Graduate step: dε_mod/dt = 0 gives e^(−t/a)(1/a + 1/b) = 1/b, so t* = a ln(1 + b/a) = 2.5 ln 3.8 = 3.34 cm, the same for both gases. At t* the moderated efficiency is 0.3853 for BF₃ (1 atm, 20 cm) and 0.4257 for He-3 (4 atm, 5 cm).
Table D: Calibration Survey with a 1.0 Ci AmBe Source Upper undergraduateGraduate step
| Room | r (cm) | Measured (cps) | Source-only (cps) | Room return (cps) | r where B = source (cm) | Fitted B (cps) |
|---|---|---|---|---|---|---|
| Small concrete room | 100 | 117.50 | 105.04 | 12.46 | 296 | 11.92 |
| Small concrete room | 300 | 23.44 | 11.67 | 11.77 | 296 | 11.92 |
| Small concrete room | 500 | 16.24 | 4.20 | 12.04 | 296 | 11.92 |
| Large hall | 300 | 15.55 | 11.67 | 3.88 | 512 | 3.81 |
| Outdoor range | 300 | 12.17 | 11.67 | 0.50 | 1449 | 0.32 |
Source term SAε/4π = 2.2 × 10⁶ × 50 × 0.12/(4π) = 1.050 × 10⁶ counts·cm²/s. Shadow-cone room return: 12.0, 4.0 and 0.5 cps.
Graduate step: the least-squares fit of C against 1/r² gave B = 11.92, 3.81 and 0.32 cps, with slopes A within 0.7–1.1% of the source term.
Discussion
The source bench confirmed the inverse-square law. The Cf-252 source lost half its output in one half-life, so its certificate must be decay-corrected before every calibration, while the AmBe and PuBe yields were effectively constant. Each 60-second count agreed with the prediction within its counting uncertainty (about √N).
Efficiency rose with pressure and length, but not in proportion, because 1 − e^(−ΣL) saturates. At 1 eV the cross section was √40 = 6.3 times smaller, so both tubes lost most of their efficiency; this is why fast-neutron counters need a moderator. He-3 at 4 atm reached 93% in only 5 cm, while BF₃ needed 20 cm to reach 84%. The wall-effect edges sat where momentum sharing predicts, so the discriminator can be set below 0.84 MeV (BF₃) or 0.19 MeV (He-3) without losing neutron counts.
The efficiency graph showed the 1/v law directly: both tubes held their thermal efficiency only up to a fraction of an electronvolt, and a denser or longer tube pushed the half-efficiency energy higher (0.29 eV for 1 atm, 20 cm BF₃ but 1.6 eV at 2 atm, 30 cm). A moderator recovers fast-neutron response, but only up to a point, because a thick moderator also absorbs and scatters away the neutrons it slows down. The distance survey showed that room return is almost constant across the room; it is negligible at 100 cm but makes up half of the signal at about 3 m in a small concrete room.
The graduate steps turned these observations into design and calibration numbers. Setting dε_mod/dt = 0 gave an optimum of 3.34 cm of polyethylene for this simplified model, independent of the gas, which is why practical moderated counters are built around a few centimetres of moderator. Fitting C = A/r² + B separated the source signal from the room return without a shadow cone; the fitted B agreed with the shadow-cone values to within a few tenths of a count per second, and the fitted slope recovered the source term to about 1%.
Conclusion
Neutron sources are calibrated by emission rate and corrected for decay and distance, and gas detectors count neutrons with an efficiency set by the absorber density, cross section, and path length.
Practice Questions
Show all work and include units.