Theory: Radiation Detection, Shielding, and Industrial Sources
Where You Will Meet Radiation at Work
Sealed radioactive sources are everyday tools in industry. Fixed gauges measure the thickness of steel, aluminum foil, paper, and plastic film on production lines, the fill level of cans and tanks, and the density of soil and asphalt. Industrial radiography uses intense gamma sources to inspect welds in pipelines, pressure vessels, and aircraft parts without cutting them open. Nuclear power plants, hospitals, and research reactors all employ radiation protection technicians who check detectors, survey work areas, and set up boundaries. In this lab you work as a technician at a manufacturing plant: you check a detector, characterize a source, design shielding, run a production-line thickness gauge, and set the safety boundary for a radiography job.
Types of Ionizing Radiation
Unstable nuclei decay by emitting radiation. The three common types differ greatly in how far they travel through matter.
| Radiation | What it is | Example source | Stopped by |
|---|---|---|---|
| Alpha (α) | Helium nucleus, charge +2, heavy | Am-241 (5.49 MeV) | A sheet of paper, or about 4 cm of air |
| Beta (β⁻) | Fast electron from the nucleus | Sr-90/Y-90 (2.28 MeV maximum) | A few millimetres of aluminum or plastic |
| Gamma (γ) | High-energy photon, no charge | Cs-137 (662 keV), Ir-192 (about 380 keV average) | Never stopped completely; reduced exponentially by dense materials such as lead |
Activity and Decay
The activity of a source is the number of decays per second, measured in becquerels (1 Bq = 1 decay/s). Industry still often uses the curie (1 Ci = 3.7 × 10¹⁰ Bq = 37 GBq). Activity falls with time as A = A₀e^(−λt), where λ = ln 2 / t½. Gauges are recalibrated regularly because their sources slowly decay: Cs-137 has a half-life of 30.1 years, Sr-90 28.8 years, and Ir-192 only 73.8 days, which is why radiography sources must be replaced several times a year.
The Geiger-Müller Detector
A Geiger-Müller (GM) tube is a gas-filled tube with a thin central wire held at several hundred volts. Radiation that enters ionizes the gas, and the freed electrons accelerate toward the wire, creating an avalanche that produces one electrical pulse. A scaler counts the pulses. The tube is operated on its plateau, the voltage range where the count rate barely changes with voltage. After each pulse the tube is "dead" for roughly 100 μs, so at very high count rates some events are lost (dead time). A GM tube counts events but cannot measure their energy.
Counting Statistics
Radioactive decay is random, so repeated counts scatter. Counts follow a Poisson distribution, and the standard deviation of a count N is simply its square root. A larger count gives a smaller relative uncertainty, which is why technicians count longer for precise work. Every measurement includes background radiation from the surroundings, which must be measured separately and subtracted.
Count rate R = N / t (counts per second, cps)
Net rate R_net = N_gross / t_gross − N_bkg / t_bkg
σ(R_net) = √( N_gross / t_gross² + N_bkg / t_bkg² )
Distance: the Inverse-Square Law
Radiation from a small (point) source spreads out over the surface of a sphere, so its intensity falls with the square of the distance. Doubling your distance from a source cuts the dose rate to one quarter. Distance is the cheapest and most powerful protection a worker has. The law holds well once the distance is several times larger than the source and the detector window; very close to the detector it breaks down.
Shielding: Exponential Attenuation and the Half-Value Layer
A narrow beam of gamma rays passing through an absorber of thickness x is reduced exponentially. The linear attenuation coefficient μ depends on the photon energy and the material, and is largest for dense, high atomic number materials such as lead and tungsten. The half-value layer (HVL) is the thickness that cuts the intensity in half. Beta particles from a source with a spread of energies are also absorbed in an approximately exponential way over thin absorbers, which is the principle of the beta thickness gauge.
Solving for thickness: x = ln(I₀ / I) / μ
Mass attenuation coefficient: μ/ρ (cm²/g), so μ = (μ/ρ) × ρ
| Radiation and absorber | Value used in this lab | Notes |
|---|---|---|
| Cs-137 γ (662 keV) in lead | μ = 1.25 cm⁻¹, HVL ≈ 0.55 cm | Narrow-beam value |
| Cs-137 γ in aluminum | μ = 0.20 cm⁻¹ | Light metals shield gamma poorly |
| Sr-90/Y-90 β in aluminum | μ/ρ = 6.6 cm²/g, μ = 17.8 cm⁻¹ (ρ = 2.70 g/cm³) | Basis of the foil thickness gauge |
| Ir-192 γ in lead | HVL = 0.48 cm | Practical value for radiography shielding |
Dose Rate, Limits, and Radiography Boundaries
Biological effect is described by dose in sieverts (Sv). For a gamma point source, the dose rate at distance r is found from the source's dose-rate constant Γ. In the United States, the occupational limit is 50 mSv per year, and an area where a person could receive more than 0.02 mSv (20 μSv) in any one hour must be controlled. During radiography, the crew ropes off a boundary where the dose rate falls to 20 μSv/h and posts warning signs. Every exposure is also kept As Low As Reasonably Achievable (ALARA) using the three tools of time, distance, and shielding.
Γ(Ir-192) = 0.13 mSv·m² / (h·GBq), A in GBq, r in m
Boundary radius where Ḋ = 0.020 mSv/h: r = √( Γ A T / 0.020 ), with T = (1/2)^(x / HVL)
Dose received: D = Ḋ × t
Time
Dose = dose rate × time. Plan the job, rehearse it, and work quickly near the source.
Distance
Dose rate ∝ 1/r². Tongs, remote cranks, and long guide tubes keep hands and bodies far away.
Shielding
Paper for alpha, aluminum or plastic for beta, lead or tungsten for gamma. Each HVL halves the dose rate.
Planning a Radiography Job with Graphs
Radiography crews plan each job before the source leaves its camera. On a log–log plot of dose rate against distance, the inverse-square law is a straight line, so the boundary is simply where that line crosses the 0.020 mSv/h limit; a collimator shifts the whole line down by its transmission. Sources also decay: with a 73.83-day half-life, an Ir-192 source loses about 1% of its activity per day. Crews read the activity on the job date from a decay curve and lengthen the exposure so that activity × time, and so the film exposure, stays the same. Sources are usually exchanged when they fall to a few hundred gigabecquerels, because exposure times become too long.
Apparatus
The equipment a radiation technician uses on this job. In the simulation these are modelled for you, and the readings correspond to what each instrument would show in a real plant.
Instructions: Running the Virtual Experiment
You are the radiation protection technician at a manufacturing plant. The simulation has three immersive 3D work areas and two graph work areas. The parts build from undergraduate work in Parts A to E to graduate-level analysis in the last step of Parts F and G; the level of each part is shown beside its title. In each one you set up the equipment, take readings, work out the answer yourself, enter it, and press Check. Record every reading and every prediction in your lab report tables, with a labelled screenshot of each setup. Drag in any 3D view to orbit, and scroll to zoom.
Simulation: Radiation Technician Work Areas and Job-Planning Graphs
A GM tube and scaler sit on a distance rail. Choose a sealed source, set the source-to-detector distance and an absorber, then count. Coloured tracks show the radiation leaving the source: red for alpha, blue for beta, and yellow for gamma. Tracks that end at the absorber were stopped. Parts A, B, and C use this bench.
Drag to orbit · scroll to zoom. 3D counting bench: a GM tube and scaler on a distance rail with a source holder, absorber plate, tongs and a lead storage pig.
Setup
Scaler and certificate
Work it out, then check
| Source | r (cm) | Absorber | t (s) | Gross N | σ = √N | Gross rate (cps) | Net rate (cps) |
|---|
An aluminum foil line passes through a C-frame beta gauge: a Sr-90 source below the foil and a detector above it. Thicker foil absorbs more beta particles, so the count rate falls. Calibrate with the gap empty, then inspect each sheet as it arrives. Each reading is a 10 s count. Part D uses this line. Use μ = 17.8 cm⁻¹ for these beta particles in aluminum.
Drag to orbit · scroll to zoom. 3D production line: an aluminum foil conveyor passing through a C-frame beta gauge, with the control cabinet display and pass/fail lamp.
Gauge controls
Specification: 0.50 ± 0.05 mm. Background rate: 0.5 cps.
Gauge display
Work it out, then check
| Sheet | R₀ (cps) | R (cps) | Your x (mm) | Measured x (mm) | Decision |
|---|
A pipeline weld is being radiographed in an open yard. The yellow camera cranks the Ir-192 source out through the guide tube to the weld. You must rope off the area where the dose rate exceeds 0.020 mSv/h and check the dose to the radiographer at the crank. Distances in the 3D view are not to scale. Part E uses this site.
Drag to orbit · scroll to zoom. 3D radiography yard: a pipe weld with the Ir-192 source, the yellow camera and guide tube, the radiographer, and the rope barrier with warning posts.
Job setup
Survey meter at the rope
Γ(Ir-192) = 0.13 mSv·m²/(h·GBq); HVL of lead = 0.48 cm.
Work it out, then check
| A (GBq) | Lead (mm) | Your boundary r (m) | Correct r (m) | d (m) | t (min) | Worker dose (mSv) |
|---|
Before a radiography job the boundary is planned on paper. The dose rate falls with the square of the distance, and a lead collimator around the source cuts it further. Part F uses this work area.
Dose rate against distance from an Ir-192 radiography source (logarithmic axes), bare and with the lead collimator you chose, with the 0.020 mSv/h public boundary line. The dots mark where each line crosses the limit; work out the exact boundary distance and check it.
Show the plotted values
Setup
Readings
Dose rate = ΓA·T/r², with T = 0.5^(x/HVL). The boundary is where the dose rate equals 0.020 mSv/h.
Work it out, then check
| A (GBq) | Collimator (mm Pb) | Boundary, bare (m) | Boundary, collimated (m) | Dose rate at 50 m (mSv/h) |
|---|
Ir-192 decays with a 73.83-day half-life, so a source loses about 1% of its activity each day. The exposure time for each film must be increased as the source decays. Part G uses this work area.
Ir-192 source activity against days since the calibration date, with the typical 370 GBq replacement level. The dot is the job date you chose.
Show the plotted values
Setup
Readings
A = A₀ × 0.5^(t/73.83 d). Exposure time × activity is kept constant: t = t₀ × A₀/A.
Work it out, then check
| A₀ (GBq) | Days | Activity (GBq) | t₀ (min) | Exposure time (min) | Days to 370 GBq |
|---|
These immersive 3D work areas and interactive graph work areas are the main version; the accessible copy delivers the same set-up, work-out, and check exercises with native controls and screen-reader announcements.
Team Questions
Example Lab Report
Sample report showing the expected format and level of detail. Use it as a guide for your own submission, and include labelled screenshots of each setup and reading. Your counts will differ slightly because radioactive decay is random.
Radiation Detection, Shielding, and Industrial Source Safety
Nuclear Engineering Technology | Section: [Your Section] | Date: [Date]
Lab Members: [Names of all members present]
Purpose
To check a Geiger-Müller counting system and apply counting statistics; to verify the inverse-square law and exponential attenuation for gamma radiation; to compare the penetration of alpha, beta, and gamma radiation; and to apply these principles to two industrial tasks: inspecting aluminum foil with a beta thickness gauge, and setting the public dose-rate boundary for an Ir-192 radiography job, including planning the radiography boundary from a dose-rate graph and correcting the source activity and exposure time for decay.
Theory
Counts follow Poisson statistics, so σ = √N. The net rate is the gross rate minus the background rate. A point source obeys I₂ = I₁(r₁/r₂)², and a narrow beam is attenuated as I = I₀e^(−μx), with HVL = 0.693/μ. A beta gauge gives thickness from x = ln(R₀/R)/μ. For radiography, Ḋ = ΓA/r² × (1/2)^(x/HVL), and the boundary is where Ḋ = 0.020 mSv/h.
Table A: Background and Counting Statistics (t = 60 s) Undergraduate
| Count | Source | N | σ = √N | Rate (cps) |
|---|---|---|---|---|
| 1 | Background | 28 | 5.3 | 0.47 |
| 2 | Background | 33 | 5.7 | 0.55 |
| 3 | Background | 29 | 5.4 | 0.48 |
| 4 | Cs-137 at 5.0 cm | 4836 | 69.5 | 80.6 |
Mean background = 0.50 cps. Net rate = 80.6 − 0.50 = 80.1 ± 1.2 cps, in agreement with the certificate value of 80 cps.
Table B: Inverse-Square Law (Cs-137, no absorber) Undergraduate
| r (cm) | Predicted net rate (cps) | Gross N in 60 s | Measured net rate (cps) |
|---|---|---|---|
| 3.0 | 222 | 13 391 | 222.7 |
| 5.0 | 80.0 | 4836 | 80.1 |
| 7.0 | 40.8 | 2465 | 40.6 |
| 10.0 | 20.0 | 1229 | 20.0 |
| 15.0 | 8.9 | 565 | 8.9 |
Sample calculation at 10 cm: 80.0 × (5.0/10.0)² = 20.0 cps. Am-241 alpha counts fell to background beyond about 4 cm, the range of 5.5 MeV alpha particles in air.
Table C1: Penetration (r = 5.0 cm for beta and gamma, 2.0 cm for alpha) Undergraduate
| Source | 1 sheet paper | 3 mm aluminum | 5 mm lead |
|---|---|---|---|
| Am-241 (α) | Stopped | Stopped | Stopped |
| Sr-90 (β) | Passes (95%) | Stopped (0.5%) | Stopped |
| Cs-137 (γ) | Passes | Passes (94%) | Passes (54%) |
Table C2: Lead Attenuation of Cs-137 (r = 5.0 cm, μ = 1.25 cm⁻¹) Undergraduate
| Lead x (mm) | Predicted net rate (cps) | Measured net rate (cps) | Fraction of unshielded |
|---|---|---|---|
| 0 | 80.0 | 80.1 | 1.00 |
| 5.5 | 40.2 | 40.5 | 0.50 |
| 11.0 | 20.2 | 19.9 | 0.25 |
| 16.5 | 10.2 | 10.3 | 0.13 |
| 22.0 | 5.1 | 5.2 | 0.06 |
HVL = 0.693/1.25 = 0.554 cm. Each 5.5 mm of lead halved the count rate, as predicted.
Table D: Foil Thickness Gauge (μ = 17.8 cm⁻¹, R₀ = 4000.6 cps gross) Undergraduate
| Sheet | R gross (cps) | x = ln[(R₀ − 0.5)/(R − 0.5)]/μ (mm) | Decision (0.50 ± 0.05 mm) |
|---|---|---|---|
| 1 | 1642.9 | 0.500 | Accept |
| 2 | 1796.0 | 0.450 | Accept (at limit) |
| 3 | 1476.7 | 0.560 | Reject |
| 4 | 1701.2 | 0.480 | Accept |
| 5 | 1520.4 | 0.544 | Accept |
Table E: Radiography Boundary (Ir-192, Γ = 0.13 mSv·m²/(h·GBq), HVL Pb = 0.48 cm) Undergraduate
| A (GBq) | Lead (mm) | Boundary r (m) | Worker d (m) | t (min) | Worker dose (mSv) |
|---|---|---|---|---|---|
| 1850 | 0 | 109.7 | 25 | 5 | 0.032 |
| 1850 | 9.6 | 54.8 | 25 | 5 | 0.0080 |
| 925 | 0 | 77.5 | 25 | 5 | 0.016 |
| 1850 | 0 | 109.7 | 50 | 5 | 0.0080 |
Sample calculation, row 1: Ḋ(1 m) = 0.13 × 1850 = 240.5 mSv/h; r = √(240.5/0.020) = 109.7 m. At 25 m, Ḋ = 240.5/625 = 0.385 mSv/h, so 5 min gives 0.032 mSv.
Table F: Radiography Boundary Planning (Graph: Boundary Distance) Upper undergraduateGraduate step
| A (GBq) | Collimator (mm Pb) | Boundary, bare (m) | Boundary, collimated (m) | Dose rate at 50 m (mSv/h) |
|---|---|---|---|---|
| 1850 | 0.0 | 109.7 | 109.7 | 0.0962 |
| 1850 | 4.8 | 109.7 | 77.5 | 0.0481 |
| 1850 | 9.6 | 109.7 | 54.8 | 0.0240 |
| 1850 | 14.4 | 109.7 | 38.8 | 0.0120 |
| 2960 | 9.6 | 138.7 | 69.4 | 0.0385 |
| 1110 | 9.6 | 84.9 | 42.5 | 0.0144 |
Graduate step: with broad-beam build-up (B = 1.20, 1.35 and 1.48 for 4.8, 9.6 and 14.4 mm) the collimated boundary at 1850 GBq becomes 84.9, 63.7 and 47.2 m instead of 77.5, 54.8 and 38.8 m; at 9.6 mm it is 80.6 m for 2960 GBq and 49.3 m for 1110 GBq.
r = √(ΓA·T/0.020), T = 0.5^(x/4.8 mm). Include a screenshot of the dose-rate graph with the boundary marked.
Table G: Ir-192 Source Decay (Graph: Source Decay) Upper undergraduateGraduate step
| A₀ (GBq) | Days | Activity (GBq) | t₀ (min) | Exposure time (min) | Days to 370 GBq |
|---|---|---|---|---|---|
| 1850 | 0 | 1850.0 | 3 | 3.00 | 171.4 |
| 1850 | 30 | 1395.9 | 3 | 3.98 | 171.4 |
| 1850 | 74 | 923.5 | 3 | 6.01 | 171.4 |
| 1850 | 120 | 599.6 | 3 | 9.26 | 171.4 |
| 1850 | 180 | 341.4 | 3 | 16.26 | 171.4 |
| 1850 | 240 | 194.4 | 3 | 28.56 | 171.4 |
Graduate step: averaging t₀·2^(t/73.83) from calibration to 370 GBq gives t_avg = t₀ × 73.83/(ln2 × D) × (A₀/370 − 1): 7.46 min for 1850 GBq (D = 171.4 d) and 10.10 min for 2960 GBq (D = 221.5 d), with t₀ = 3.0 min.
A = A₀ × 0.5^(t/73.83 d); t = t₀ × A₀/A. A 1850 GBq source reaches the 370 GBq replacement level after 171.4 days.
Discussion
The background was steady at about 0.5 cps, and the scatter between repeated counts matched σ = √N, confirming the counter was working correctly. The Cs-137 net rate fell as 1/r², and each 5.5 mm of lead halved it, in agreement with μ = 1.25 cm⁻¹. The penetration test showed the expected pattern: paper stopped alpha, a few millimetres of aluminum stopped beta, and gamma rays were only reduced, even by lead. Aluminum is preferred for beta shielding because a high atomic number material such as lead turns more of the beta energy into bremsstrahlung X-rays.
The foil gauge converted count rates to thickness with a statistical uncertainty of about 0.005 mm, more than enough to sort sheets against a ±0.05 mm specification. In the radiography job, two half-value layers of lead (9.6 mm) cut the boundary radius in half, and doubling the radiographer's distance cut the radiographer's dose by a factor of four, showing how shielding, distance, and time together keep both the public and the crew within limits.
The graphs turned these results into job-planning tools. On the log–log dose-rate graph each collimator thickness gave a parallel straight line, and every half-value layer moved the boundary inward by a factor of √2: from 109.7 m bare to 77.5, 54.8 and 38.8 m for one, two and three HVLs at 1850 GBq. The decay curve showed why the exposure time must be updated on every job: four months after calibration the source had fallen to 600 GBq and each film needed 9.3 min instead of 3 min. The radiographer also gains from the decay, because the same boundary rule now gives a smaller roped area.
The graduate steps showed where the simple models fall short. Narrow-beam attenuation underestimated the dose behind the collimator: with build-up the 9.6 mm boundary moved from 54.8 m to 63.7 m, so a rope set from the half-value layer alone would leave the public above 0.020 mSv/h. Averaged over its working life, a 1850 GBq source needed 7.46 min per film, about 2.5 times the fresh-source time; a larger 2960 GBq source lasts longer but its average time rises to 10.1 min, which adds crew time near the camera and must be weighed in the ALARA plan.
Conclusion
The counting system obeyed Poisson statistics, gamma intensity followed the inverse-square law and exponential attenuation with an HVL of 0.55 cm of lead, and the three radiation types showed their characteristic penetration. These principles allowed accurate thickness measurement on a production line and a correctly placed radiography boundary of 109.7 m for an unshielded 1850 GBq Ir-192 source.
Practice Questions
Show all work and include units. Use the equations and data in the Theory section.