Theory: Water Radiolysis and Chemical Dosimetry
Why Radiation Chemistry Matters
Ionizing radiation does chemistry. In reactor coolant it splits water into hydrogen, oxygen, and hydrogen peroxide, which drive corrosion and must be controlled with dissolved hydrogen. In sealed waste drums it generates flammable hydrogen gas, so shipping rules limit the gas build-up. In medical and industrial irradiators the chemical change is used to measure dose. Radiation chemists in power plants, sterilization facilities, and waste programs use the same few quantities: the absorbed dose and the radiation-chemical yield.
Absorbed Dose and G-Values
The absorbed dose D is the energy deposited per unit mass, in grays (1 Gy = 1 J/kg). The radiation-chemical yield G is the number of molecules formed or destroyed per 100 eV of energy absorbed. In SI units it is given in µmol/J.
Amount formed: n (mol/kg) = G (mol/J) × D (Gy)
The Stages of Water Radiolysis
Within 10⁻¹⁵ s radiation ionizes and excites water molecules along its track. The fragments react inside small clusters called spurs. By about 10⁻⁷ s the spurs have spread out and the primary products remain: the hydrated electron e⁻(aq), the hydrogen atom H·, the hydroxyl radical ·OH, molecular hydrogen H₂, and hydrogen peroxide H₂O₂.
| Species | Gamma or fast electrons (low LET) | Alpha particles (high LET) |
|---|---|---|
| e⁻(aq) | 2.6 | 0.15 |
| H· | 0.6 | 0.20 |
| ·OH | 2.7 | 0.35 |
| H₂ | 0.45 | 1.30 |
| H₂O₂ | 0.7 | 1.00 |
| HO₂· | ≈ 0 | 0.20 |
Values are in molecules per 100 eV and are representative of neutral water; published values vary slightly. Linear energy transfer (LET) is the energy lost per unit path length. A gamma ray leaves spurs far apart, so most radicals escape. An alpha particle leaves a dense column of overlapping spurs, so radicals recombine and more molecular H₂ and H₂O₂ form. Because water is split into equal oxidizing and reducing equivalents, the yields must balance:
The Fricke Dosimeter
The Fricke dosimeter is 1 mM ferrous ammonium sulphate and 1 mM NaCl in air-saturated 0.4 M sulphuric acid. The radiolysis products oxidize Fe²⁺ to Fe³⁺. In acid, e⁻(aq) becomes an H atom. Each ·OH oxidizes one ion, each H₂O₂ two, and each H atom (through HO₂·) three, so G(Fe³⁺) = 3G(H) + G(·OH) + 2G(H₂O₂). For cobalt-60 gamma rays G(Fe³⁺) = 15.5 ions/100 eV (1.607 µmol/J). Fe³⁺ absorbs at 304 nm, so the dose is found with the Beer–Lambert law.
ε(25 °C) = 2174 L mol⁻¹ cm⁻¹; ε(T) = ε(25 °C) [1 + 0.007 (T − 25)]; ρ = 1.024 kg/L
Irradiation facilities use it to calibrate the dose rate at each position in a gamma cell. The source decays (Co-60 half-life 5.27 years), so the calibration is repeated and the dose rate is corrected for decay between calibrations.
Calibration Fits and Hydrogen Mass Balance
A batch of dosimeter solution is calibrated by plotting ΔA against known doses and fitting a line through the origin. The slope b = εlρG gives the G-value directly, and the scatter of the points about the line gives its uncertainty. In a sealed waste drum the hydrogen made each day, V = 22.414 G Ḋ m, collects in the headspace, so the fraction rises linearly until it reaches the 4% lower flammability limit. A vent that exchanges Q litres of headspace gas per day gives V_h dc/dt = V − Qc, which settles at c = V/Q.
Apparatus
The equipment used for chemical dosimetry and radiolysis studies. In the simulation these are modelled for you.
Instructions: Running the Virtual Experiment
You are the radiation chemist on duty. The simulation has two immersive 3D work areas and two graph work areas. The parts build from undergraduate work in Parts A and B to graduate-level analysis in the last step of Parts C and D; the level of each part is shown beside its title. In each one you set the conditions, work out the answer yourself, enter it, and press Check. Record every reading and calculation in your lab report tables (modelled in the Example Report), with a labelled screenshot of each setup. Drag in any 3D view to orbit, and scroll to zoom.
Simulation: Gamma Cell, Radiolysis Bench, and Analysis Graphs
The Fricke dosimeter is a solution of ferrous sulphate in dilute sulphuric acid. Radiation oxidizes Fe²⁺ to Fe³⁺, and the Fe³⁺ absorbs strongly at 304 nm. It is the classic reference chemical dosimeter for calibrating gamma irradiators. Part A uses this work area.
Setup
Readings
ΔA = ε(T) · l · ρ · G · D, with G in mol/J and D in Gy (J/kg). Above about 400 Gy the dissolved oxygen is used up and the response is no longer linear.
Work it out, then check
| Position | Time (min) | Nominal D (Gy) | T (°C) | Predicted ΔA | Measured ΔA | Measured D (Gy) | Measured rate (Gy/min) |
|---|
When radiation passes through water it leaves small clusters of ions and excited molecules called spurs. Within about a microsecond these form the primary radiolysis products. Their yields are expressed as G-values. Part B uses this work area.
Setup
Readings
n (mol/kg) = G (molecules/100 eV) × 1.036 × 10⁻⁷ × D (Gy). These are representative G-values for teaching.
Work it out, then check
| Radiation | Species | G (per 100 eV) | G (µmol/J) | Dose (kGy) | n (mol/kg) | Water (kg) | H₂ (L at STP) |
|---|
A dosimetry laboratory calibrates each new batch of Fricke solution by irradiating vials to known doses and fitting a line through the readings. The slope then converts any later reading into dose. Part C uses this work area.
Setup
Readings
Fit through the origin: slope b = Σ(D·ΔA)/ΣD². Calibration data are simplified training data.
Work it out, then check
| Unknown vial | ΔA | Slope b (per Gy) | Dose (Gy) | G (µmol/J) | u(G)/G (%) |
|---|
Waste that contains water and radioactivity keeps making hydrogen. In a sealed drum the gas collects in the headspace until it could burn, so storage times and vents are set from the generation rate. Part D uses this work area.
Setup
Readings
n = G × D_rate × m (mol/day); V = 22.414 n L/day; sealed fraction c(t) = V t / V_h. Simplified training model: no leakage, no recombination.
Work it out, then check
| Radiation | Dose rate (Gy/day) | Water (kg) | H₂ (mL/day) | Days to 4% | Vent target | Vent rate (L/day) |
|---|
This accessible version provides the same set-up, work-out, and check exercises as the 3D and graph work areas, using native controls, data tables, and screen-reader announcements.
Team Questions
Example Lab Report
Sample report showing the expected format and level of detail. Use it as a guide for your own submission, and include labelled screenshots of each setup. Each measurement carries a small random reading error, so your measured values will differ slightly.
Radiation Chemistry: Water Radiolysis and Fricke Dosimetry
Nuclear Chemistry | Section: [Your Section] | Date: [Date]
Lab Members: [Names of all members present]
Purpose
To calibrate the dose rate at three positions in a cobalt-60 gamma cell with the Fricke dosimeter, to use radiolysis G-values to calculate product yields and hydrogen generation for gamma and alpha radiation, to calibrate the dosimeter by least squares and recover G(Fe³⁺) with its uncertainty, and to predict hydrogen build-up in sealed waste drums and size a vent.
Theory
The Fricke response is linear in dose up to about 400 Gy. G-values convert absorbed energy into moles of product.
Table A: Fricke Dosimetry Undergraduate
| Position | Time (min) | Nominal D (Gy) | T (°C) | Predicted ΔA | Measured ΔA | Measured D (Gy) | Measured rate (Gy/min) |
|---|---|---|---|---|---|---|---|
| Centre | 10 | 80.0 | 25 | 0.2862 | 0.2851 | 79.7 | 7.969 |
| Mid-radius | 10 | 64.0 | 25 | 0.2290 | 0.2296 | 64.2 | 6.418 |
| Outer edge | 10 | 48.0 | 25 | 0.1717 | 0.1712 | 47.9 | 4.786 |
| Centre | 25 | 200.0 | 30 | 0.7405 | 0.7431 | 200.7 | 8.028 |
| Centre | 60 | 480.0 | 25 | saturated | 1.5309 | — | — |
Table B: Radiolysis Yields (10 kGy, 200 kg of water) Undergraduate
| Radiation | Species | G (per 100 eV) | G (µmol/J) | Dose (kGy) | n (mol/kg) | Water (kg) | H₂ (L at STP) |
|---|---|---|---|---|---|---|---|
| Gamma | H₂ | 0.45 | 0.0466 | 10 | 4.664 × 10⁻⁴ | 200 | 2.09 |
| Gamma | ·OH | 2.7 | 0.2798 | 10 | 2.798 × 10⁻³ | 200 | 2.09 |
| Gamma | H₂O₂ | 0.7 | 0.0725 | 10 | 7.255 × 10⁻⁴ | 200 | 2.09 |
| Alpha | H₂ | 1.30 | 0.1347 | 10 | 1.347 × 10⁻³ | 200 | 6.04 |
| Alpha | H₂O₂ | 1.00 | 0.1036 | 10 | 1.036 × 10⁻³ | 200 | 6.04 |
Material balance: gamma 2.7 + 2(0.7) = 4.10 = 2.6 + 0.6 + 2(0.45); alpha 0.35 + 2(1.00) + 3(0.20) = 2.95 = 0.15 + 0.20 + 2(1.30).
Table C: Least-Squares Calibration Upper undergraduateGraduate step
| Unknown vial | ΔA | Slope b (per Gy) | Dose (Gy) | G (µmol/J) | u(G)/G (%) |
|---|---|---|---|---|---|
| U-1 | 0.4210 | 3.5717 × 10⁻³ | 117.9 | 1.6044 | 0.122 |
| U-2 | 0.8650 | 3.5717 × 10⁻³ | 242.2 | 1.6044 | 0.122 |
| U-3 | 1.2200 | 3.5717 × 10⁻³ | 341.6 | 1.6044 | 0.122 |
b = Σ(D·ΔA)/ΣD² = 1252.325/350625 = 3.5717 × 10⁻³ per Gy; G = b/(ε l ρ) = 1.6044 µmol/J, 0.2% below the accepted 1.607.
Graduate step: Σ residual² = 4.67 × 10⁻⁵, u(b) = 4.36 × 10⁻⁶ per Gy, so u(G)/G = 0.122% and G = 1.6044 ± 0.0020 µmol/J.
Table D: Hydrogen in Sealed Drums (50 L headspace) Upper undergraduateGraduate step
| Radiation | Dose rate (Gy/day) | Water (kg) | H₂ (mL/day) | Days to 4% | Vent target | Vent rate (L/day) |
|---|---|---|---|---|---|---|
| Alpha | 50 | 200 | 30.20 | 66.2 | 50% LFL | 1.510 |
| Alpha | 20 | 100 | 6.04 | 331.1 | 50% LFL | 0.302 |
| Gamma | 50 | 200 | 10.45 | 191.3 | 50% LFL | 0.523 |
| Gamma | 20 | 100 | 2.09 | 956.6 | 50% LFL | 0.105 |
| Alpha | 50 | 200 | 30.20 | 66.2 | 25% LFL | 3.020 |
Graduate step: steady state c = V/Q. For the alpha drum at 50 Gy/day and 200 kg, Q = 1.510 L/day holds 2% H₂ and 3.020 L/day holds 1%; the time constants V_h/Q are 33 and 17 days.
Discussion
The measured dose rates agreed with the nominal calibration within 0.5%, and fell from 7.97 Gy/min at the centre to 4.79 Gy/min at the outer edge, so samples must be placed at the calibrated position. Reading at 30 °C raised ε by 3.5%; without that correction the dose would have been overstated by the same amount. The 60-minute vial (480 Gy) was saturated because the dissolved oxygen had been used up, so it could not be used; vials must stay below about 400 Gy.
Alpha radiation gave nearly three times as much H₂ as gamma radiation for the same dose, because its dense track lets radicals recombine into molecular products before they escape. Waste containing alpha emitters and water therefore needs vented lids or limits on the gas build-up, and reactor coolant chemistry adds dissolved H₂ to suppress the net radiolysis caused by the radicals.
The least-squares calibration turned eight readings into a slope of 3.5717 × 10⁻³ per Gy and a G-value of 1.6044 µmol/J, within 0.2% of the accepted value. The fit itself contributes only about 0.1% uncertainty, so a real calibration is limited by the molar absorption coefficient, the cell temperature and the reference dose. The calibrated slope dated the unknown vials at 118, 242 and 342 Gy.
The drum graph showed the practical cost of a high G(H₂). The alpha drum reached the 4% flammability limit in 66 days, against 191 days for the same gamma dose, which is why alpha-bearing waste is shipped in vented drums. The mass balance at the graduate step gave a vent rate of about 1.5 L/day to hold the headspace at half the limit, a flow that a small sintered-carbon filter vent can pass.
Conclusion
Radiation-chemical yields link absorbed energy directly to chemical change. The Fricke dosimeter uses this to measure dose, and the same G-values predict the gas and peroxide generated in reactors and waste packages.
Practice Questions
Show all work and include units.