Theory: Water Radiolysis and Chemical Dosimetry

Why Radiation Chemistry Matters

Ionizing radiation does chemistry. In reactor coolant it splits water into hydrogen, oxygen, and hydrogen peroxide, which drive corrosion and must be controlled with dissolved hydrogen. In sealed waste drums it generates flammable hydrogen gas, so shipping rules limit the gas build-up. In medical and industrial irradiators the chemical change is used to measure dose. Radiation chemists in power plants, sterilization facilities, and waste programs use the same few quantities: the absorbed dose and the radiation-chemical yield.

Absorbed Dose and G-Values

The absorbed dose D is the energy deposited per unit mass, in grays (1 Gy = 1 J/kg). The radiation-chemical yield G is the number of molecules formed or destroyed per 100 eV of energy absorbed. In SI units it is given in µmol/J.

Yield Conversion 1 molecule/100 eV = 1.0364 × 10⁻⁷ mol/J = 0.10364 µmol/J
Amount formed: n (mol/kg) = G (mol/J) × D (Gy)
Gas volume at STP: V = n × m × 22.414 L/mol

The Stages of Water Radiolysis

Within 10⁻¹⁵ s radiation ionizes and excites water molecules along its track. The fragments react inside small clusters called spurs. By about 10⁻⁷ s the spurs have spread out and the primary products remain: the hydrated electron e⁻(aq), the hydrogen atom H·, the hydroxyl radical ·OH, molecular hydrogen H₂, and hydrogen peroxide H₂O₂.

SpeciesGamma or fast electrons (low LET)Alpha particles (high LET)
e⁻(aq)2.60.15
H·0.60.20
·OH2.70.35
H₂0.451.30
H₂O₂0.71.00
HO₂·≈ 00.20

Values are in molecules per 100 eV and are representative of neutral water; published values vary slightly. Linear energy transfer (LET) is the energy lost per unit path length. A gamma ray leaves spurs far apart, so most radicals escape. An alpha particle leaves a dense column of overlapping spurs, so radicals recombine and more molecular H₂ and H₂O₂ form. Because water is split into equal oxidizing and reducing equivalents, the yields must balance:

Material Balance G(·OH) + 2G(H₂O₂) + 3G(HO₂) = G(e⁻aq) + G(H·) + 2G(H₂)

The Fricke Dosimeter

The Fricke dosimeter is 1 mM ferrous ammonium sulphate and 1 mM NaCl in air-saturated 0.4 M sulphuric acid. The radiolysis products oxidize Fe²⁺ to Fe³⁺. In acid, e⁻(aq) becomes an H atom. Each ·OH oxidizes one ion, each H₂O₂ two, and each H atom (through HO₂·) three, so G(Fe³⁺) = 3G(H) + G(·OH) + 2G(H₂O₂). For cobalt-60 gamma rays G(Fe³⁺) = 15.5 ions/100 eV (1.607 µmol/J). Fe³⁺ absorbs at 304 nm, so the dose is found with the Beer–Lambert law.

Fricke Dose Equation ΔA = ε(T) · l · ρ · G(Fe³⁺) · D   →   D = ΔA / [ε(T) · l · ρ · G(Fe³⁺)]
ε(25 °C) = 2174 L mol⁻¹ cm⁻¹; ε(T) = ε(25 °C) [1 + 0.007 (T − 25)]; ρ = 1.024 kg/L
Useful range about 20 to 400 Gy; above that the dissolved O₂ is used up

Irradiation facilities use it to calibrate the dose rate at each position in a gamma cell. The source decays (Co-60 half-life 5.27 years), so the calibration is repeated and the dose rate is corrected for decay between calibrations.

Calibration Fits and Hydrogen Mass Balance

A batch of dosimeter solution is calibrated by plotting ΔA against known doses and fitting a line through the origin. The slope b = εlρG gives the G-value directly, and the scatter of the points about the line gives its uncertainty. In a sealed waste drum the hydrogen made each day, V = 22.414 G Ḋ m, collects in the headspace, so the fraction rises linearly until it reaches the 4% lower flammability limit. A vent that exchanges Q litres of headspace gas per day gives V_h dc/dt = V − Qc, which settles at c = V/Q.

Apparatus

The equipment used for chemical dosimetry and radiolysis studies. In the simulation these are modelled for you.

Cobalt-60 gamma cell
A lead-shielded chamber surrounded by Co-60 source pencils (1.17 and 1.33 MeV gamma rays).
Fricke dosimeter vial
Air-saturated ferrous sulphate in 0.4 M H₂SO₄, sealed in a glass ampoule.
0.2862
UV-visible spectrophotometer
Reads absorbance at 304 nm in a 1.000 cm quartz cuvette against unirradiated solution.
Thermometer
Records the solution temperature at reading, for the ε correction.
Sealed waste drum
Wet or water-bearing waste generates H₂ by radiolysis; vent filters limit the build-up.
Track structure
Isolated spurs for gamma rays; a dense column for alpha particles.

Instructions: Running the Virtual Experiment

You are the radiation chemist on duty. The simulation has two immersive 3D work areas and two graph work areas. The parts build from undergraduate work in Parts A and B to graduate-level analysis in the last step of Parts C and D; the level of each part is shown beside its title. In each one you set the conditions, work out the answer yourself, enter it, and press Check. Record every reading and calculation in your lab report tables (modelled in the Example Report), with a labelled screenshot of each setup. Drag in any 3D view to orbit, and scroll to zoom.

Part A: Calibrating a Gamma Cell with the Fricke Dosimeter (3D Gamma Cell)Undergraduate
1
Open Simulation → 3D Gamma Cell. Put the vial at the Centre, set 10 min and 25 °C. Check the dose delivered and the absorbance change ΔA.
2
Press Measure absorbance. From your measured ΔA, check the dose and the measured dose rate. Press Record this run.
3
Repeat at the Mid-radius and Outer edge positions (10 min, 25 °C), and at the Centre for 25 min read at 30 °C. Record each run in Table A.
4
Irradiate a Centre vial for 60 min and measure it. Explain in your report why the result cannot be used.
Part B: Radiolysis Yields and Hydrogen in a Waste Drum (3D Radiolysis Bench)Undergraduate
1
Open 3D Radiolysis Bench. Choose Gamma, species H₂, 10 kGy, and 200 kg of water. Check G in µmol/J, the moles per kg, and the H₂ volume in the drum. Press Record this run.
2
Repeat for ·OH and H₂O₂ with gamma radiation. Record them in Table B.
3
Change to Alpha. Repeat for H₂ and H₂O₂, and check the material balance for both radiation types. Compare the tracks in the 3D view and explain the difference in your report.
Part C: Calibrating the Dosimeter by Least Squares (Graph: Calibration Fit)Upper undergraduateGraduate step
1
Open Graph: Calibration Fit. Use the sums in the Readings panel to check the least-squares slope b of the line through the origin.
2
Choose Vial U-1 and check the dose it received. Press Record this run. Repeat for U-2 and U-3 and fill Table C.
3
From the slope, check G(Fe³⁺) in µmol/J and compare it with the accepted 1.607 µmol/J. Take a labelled screenshot of the fit.
4
GraduateUse the eight points in the data table to find the residual variance and check the relative standard uncertainty of G from the fit. Record it in Table C and discuss which other uncertainties (ε, temperature, reference dose) would dominate in a real calibration.
Part D: Hydrogen in Sealed Waste Drums (Graph: Drum Hydrogen)Upper undergraduateGraduate step
1
Open Graph: Drum Hydrogen. Choose Alpha, 50 Gy/day and 200 kg. Check the H₂ generation rate and the days for a sealed drum to reach 4%. Press Record this run.
2
Repeat for alpha at 20 Gy/day and 100 kg, and for gamma at 50 Gy/day, 200 kg and at 20 Gy/day, 100 kg. Record every run in Table D and compare the alpha and gamma curves on the graph.
3
Explain why waste-acceptance rules limit how long such drums may stay sealed.
4
GraduateWrite the headspace mass balance V_h dc/dt = V − Qc and check the steady-state vent rate that holds the alpha drum (50 Gy/day, 200 kg) at 50% and at 25% of the LFL. Record the rates in Table D and state the time constant for reaching steady state.

Simulation: Gamma Cell, Radiolysis Bench, and Analysis Graphs

Radiation Chemistry Virtual LabSet up the equipment, work out your answer, then Check

The Fricke dosimeter is a solution of ferrous sulphate in dilute sulphuric acid. Radiation oxidizes Fe²⁺ to Fe³⁺, and the Fe³⁺ absorbs strongly at 304 nm. It is the classic reference chemical dosimeter for calibrating gamma irradiators. Part A uses this work area.

About this work area: The gamma cell irradiates a vial of Fricke solution at the chosen position for the chosen time. You then read the absorbance change at 304 nm on a spectrophotometer at the chosen temperature. The Readings panel gives the dose rate at each position, the G-value, the molar absorption coefficient, and the solution density, and every change is announced.

Setup

Readings

Nominal dose rate at this position—
G(Fe³⁺)—
ε at 304 nm (25 °C)—
Temperature coefficient of ε—
Path length; solution density—
Measured ΔA at 304 nm—

ΔA = ε(T) · l · ρ · G · D, with G in mol/J and D in Gy (J/kg). Above about 400 Gy the dissolved oxygen is used up and the response is no longer linear.

Work it out, then check

Table A: your Fricke dosimetry record
PositionTime (min)Nominal D (Gy)T (°C)Predicted ΔAMeasured ΔAMeasured D (Gy)Measured rate (Gy/min)

When radiation passes through water it leaves small clusters of ions and excited molecules called spurs. Within about a microsecond these form the primary radiolysis products. Their yields are expressed as G-values. Part B uses this work area.

About this work area: The radiolysis bench lets you choose the radiation type, the absorbed dose, the mass of water in a sealed drum, and the species of interest. The Readings panel lists the G-values for that radiation, and every change is announced.

Setup

Readings

G(e⁻aq), G(H·), G(·OH)—
G(H₂), G(H₂O₂), G(HO₂)—
G of the chosen species—
Conversion—
Molar volume of gas at STP—

n (mol/kg) = G (molecules/100 eV) × 1.036 × 10⁻⁷ × D (Gy). These are representative G-values for teaching.

Work it out, then check

Table B: your radiolysis record
RadiationSpeciesG (per 100 eV)G (µmol/J)Dose (kGy)n (mol/kg)Water (kg)H₂ (L at STP)

A dosimetry laboratory calibrates each new batch of Fricke solution by irradiating vials to known doses and fitting a line through the readings. The slope then converts any later reading into dose. Part C uses this work area.

About this work area: This graph work area shows a dosimeter calibration: the measured absorbance change of eight vials against the dose each received, with a straight line fitted through the origin. You choose an unknown vial. The data table below lists every calibration point, and every change is announced.

Setup

Readings

Calibration points—
ΣD² (Gy²)—
Σ D·ΔA (Gy)—
Unknown vial ΔA—
ε, path, density—

Fit through the origin: slope b = Σ(D·ΔA)/ΣD². Calibration data are simplified training data.

Work it out, then check

Table C: your calibration record
Unknown vialΔASlope b (per Gy)Dose (Gy)G (µmol/J)u(G)/G (%)

Waste that contains water and radioactivity keeps making hydrogen. In a sealed drum the gas collects in the headspace until it could burn, so storage times and vents are set from the generation rate. Part D uses this work area.

About this work area: This graph work area plots the hydrogen fraction in the headspace of a sealed waste drum against storage time, for gamma and for alpha waste, with the 4% lower flammability limit. You choose the radiation type, the absorbed dose rate in the waste, and the mass of water. The data table below lists the curves, and every change is announced.

Setup

Readings

G(H₂)—
Headspace volume—
Lower flammability limit—
Molar volume—

n = G × D_rate × m (mol/day); V = 22.414 n L/day; sealed fraction c(t) = V t / V_h. Simplified training model: no leakage, no recombination.

Work it out, then check

Table D: your drum-hydrogen record
RadiationDose rate (Gy/day)Water (kg)H₂ (mL/day)Days to 4%Vent targetVent rate (L/day)

This accessible version provides the same set-up, work-out, and check exercises as the 3D and graph work areas, using native controls, data tables, and screen-reader announcements.

Team Questions

Question 1. A Fricke vial sits where the dose rate is 8.00 Gy/min for 12.5 min. What dose does it receive, in Gy?
Question 2. What absorbance change at 304 nm do you expect for that dose, read at 25 °C in a 1 cm cell (ε = 2174 L/mol/cm, ρ = 1.024 kg/L, G = 1.607 µmol/J)?
Question 3. Another vial gives ΔA = 0.500 read at 25 °C. What dose did it receive, in Gy?
Question 4. The same ΔA = 0.500 is read at 30 °C instead. Using ε(T) = 2174[1 + 0.007(T − 25)], what is the dose, in Gy?
Question 5. For gamma radiolysis G(·OH) = 2.7 molecules/100 eV. What is this in µmol/J?
Question 6. A sealed drum holds 200 kg of water that absorbs 10 kGy of gamma radiation. With G(H₂) = 0.45, how many litres of H₂ at STP form?
Question 7: Challenge. The same drum instead absorbs 10 kGy from alpha emitters, with G(H₂) = 1.30. How many litres of H₂ at STP form?

Example Lab Report

Sample report showing the expected format and level of detail. Use it as a guide for your own submission, and include labelled screenshots of each setup. Each measurement carries a small random reading error, so your measured values will differ slightly.

Radiation Chemistry: Water Radiolysis and Fricke Dosimetry

Nuclear Chemistry | Section: [Your Section] | Date: [Date]

Lab Members: [Names of all members present]

Purpose

To calibrate the dose rate at three positions in a cobalt-60 gamma cell with the Fricke dosimeter, to use radiolysis G-values to calculate product yields and hydrogen generation for gamma and alpha radiation, to calibrate the dosimeter by least squares and recover G(Fe³⁺) with its uncertainty, and to predict hydrogen build-up in sealed waste drums and size a vent.

Theory

The Fricke response is linear in dose up to about 400 Gy. G-values convert absorbed energy into moles of product.

D = ΔA/[ε(T) l ρ G]  ·  n = G × 1.0364×10⁻⁷ × D  ·  V = n m × 22.414 L

Table A: Fricke Dosimetry Undergraduate

PositionTime (min)Nominal D (Gy)T (°C)Predicted ΔAMeasured ΔAMeasured D (Gy)Measured rate (Gy/min)
Centre1080.0250.28620.285179.77.969
Mid-radius1064.0250.22900.229664.26.418
Outer edge1048.0250.17170.171247.94.786
Centre25200.0300.74050.7431200.78.028
Centre60480.025saturated1.5309——

Table B: Radiolysis Yields (10 kGy, 200 kg of water) Undergraduate

RadiationSpeciesG (per 100 eV)G (µmol/J)Dose (kGy)n (mol/kg)Water (kg)H₂ (L at STP)
GammaH₂0.450.0466104.664 × 10⁻⁴2002.09
Gamma·OH2.70.2798102.798 × 10⁻³2002.09
GammaH₂O₂0.70.0725107.255 × 10⁻⁴2002.09
AlphaH₂1.300.1347101.347 × 10⁻³2006.04
AlphaH₂O₂1.000.1036101.036 × 10⁻³2006.04

Material balance: gamma 2.7 + 2(0.7) = 4.10 = 2.6 + 0.6 + 2(0.45); alpha 0.35 + 2(1.00) + 3(0.20) = 2.95 = 0.15 + 0.20 + 2(1.30).

Table C: Least-Squares Calibration Upper undergraduateGraduate step

Unknown vialΔASlope b (per Gy)Dose (Gy)G (µmol/J)u(G)/G (%)
U-10.42103.5717 × 10⁻³117.91.60440.122
U-20.86503.5717 × 10⁻³242.21.60440.122
U-31.22003.5717 × 10⁻³341.61.60440.122

b = Σ(D·ΔA)/ΣD² = 1252.325/350625 = 3.5717 × 10⁻³ per Gy; G = b/(ε l ρ) = 1.6044 µmol/J, 0.2% below the accepted 1.607.

Graduate step: Σ residual² = 4.67 × 10⁻⁵, u(b) = 4.36 × 10⁻⁶ per Gy, so u(G)/G = 0.122% and G = 1.6044 ± 0.0020 µmol/J.

Table D: Hydrogen in Sealed Drums (50 L headspace) Upper undergraduateGraduate step

RadiationDose rate (Gy/day)Water (kg)H₂ (mL/day)Days to 4%Vent targetVent rate (L/day)
Alpha5020030.2066.250% LFL1.510
Alpha201006.04331.150% LFL0.302
Gamma5020010.45191.350% LFL0.523
Gamma201002.09956.650% LFL0.105
Alpha5020030.2066.225% LFL3.020

Graduate step: steady state c = V/Q. For the alpha drum at 50 Gy/day and 200 kg, Q = 1.510 L/day holds 2% H₂ and 3.020 L/day holds 1%; the time constants V_h/Q are 33 and 17 days.

Discussion

The measured dose rates agreed with the nominal calibration within 0.5%, and fell from 7.97 Gy/min at the centre to 4.79 Gy/min at the outer edge, so samples must be placed at the calibrated position. Reading at 30 °C raised ε by 3.5%; without that correction the dose would have been overstated by the same amount. The 60-minute vial (480 Gy) was saturated because the dissolved oxygen had been used up, so it could not be used; vials must stay below about 400 Gy.

Alpha radiation gave nearly three times as much H₂ as gamma radiation for the same dose, because its dense track lets radicals recombine into molecular products before they escape. Waste containing alpha emitters and water therefore needs vented lids or limits on the gas build-up, and reactor coolant chemistry adds dissolved H₂ to suppress the net radiolysis caused by the radicals.

The least-squares calibration turned eight readings into a slope of 3.5717 × 10⁻³ per Gy and a G-value of 1.6044 µmol/J, within 0.2% of the accepted value. The fit itself contributes only about 0.1% uncertainty, so a real calibration is limited by the molar absorption coefficient, the cell temperature and the reference dose. The calibrated slope dated the unknown vials at 118, 242 and 342 Gy.

The drum graph showed the practical cost of a high G(H₂). The alpha drum reached the 4% flammability limit in 66 days, against 191 days for the same gamma dose, which is why alpha-bearing waste is shipped in vented drums. The mass balance at the graduate step gave a vent rate of about 1.5 L/day to hold the headspace at half the limit, a flow that a small sintered-carbon filter vent can pass.

Conclusion

Radiation-chemical yields link absorbed energy directly to chemical change. The Fricke dosimeter uses this to measure dose, and the same G-values predict the gas and peroxide generated in reactors and waste packages.

Practice Questions

Show all work and include units.

Question 1
A Fricke vial read at 22 °C gives ΔA = 0.420 in a 1 cm cell. What dose did it receive?
Hint: ε(22 °C) = 2174 × (1 − 0.021).
Question 2
The gamma cell was calibrated at 8.00 Gy/min one year ago. Co-60 has a half-life of 5.27 years. What is the dose rate today?
Hint: multiply by e^(−ln 2 × 1/5.27).
Question 3
Show that G(Fe³⁺) = 15.5 follows from G(H) = 3.65, G(·OH) = 2.97, and G(H₂O₂) = 0.78 in 0.4 M H₂SO₄.
Hint: G(Fe³⁺) = 3G(H) + G(·OH) + 2G(H₂O₂).
Question 4
How much H₂O₂ (in mol/kg) forms in reactor coolant that absorbs 5 kGy of gamma radiation, if no radicals recombine with it?
Hint: G(H₂O₂) = 0.7.
Question 5
Why does an alpha particle give a higher G(H₂) but a lower G(·OH) than a gamma ray?
Hint: think about how close the spurs are along each track.
Question 6
Why do PWRs add dissolved hydrogen to the primary coolant?
Hint: H₂ reacts with ·OH and pushes the radiolysis back toward water.